Fluorescence Hydrogen Peroxide Probe Based on a Microstructured Polymer Optical Fiber Modified with a Titanium Dioxide Film
DONGDONG LI and LILI WANG*
A highly sensitive microstructured polymer optical fiber (MPOF) probe for hydrogen peroxide was made by forming a rhodamine 6G-doped titanium dioxide film on the side walls of array holes in an MPOF. It was found that hydrogen peroxide only has a response to the MPOF probe in a certain concentration of potassium iodide in sulfuric acid solution. The calibration graph of fluorescence intensity versus hydrogen peroxide concentration is linear in the range of 1.6 × 10-7 mol/L to 9.6 × 10-5 mol/L. The method, with high sensitivity and a wide linear range, has been applied to the determination of trace amounts of hydrogen peroxide in a few real samples, such as rain water and contact lens disinfectant, with satisfactory results.
Index Headings: Hydrogen peroxide; Fluorescence probe; Microstructured polymer optical fiber; Titanium oxide film.