holder

The following is an abstract for the selected article. A PDF download of the full text of this article is available here. Members may download full texts at no charge. Non-members may be charged a small fee for certain articles.


Time-Correlated Raman and Fluorescence Spectroscopy Based on a Silicon Photomultiplier and Time-Correlated Single Photon Counting Technique

Volume 67, Number 2 (Feb. 2013) Page 136-140

CHUNLING ZHANG, LIYING ZHANG, RU YANG, KUN LIANG, and DEJUN HAN*


We report a time-correlated Raman spectroscopy technique based on a silicon photomultiplier (SiPM) and a time-correlated single photon counting (TCSPC) technique to exploit the natural temporal separation between Raman and fluorescence phenomena to alleviate the high fluorescence background with conventional Raman detection. The TCSPC technique employed can greatly reduce the effect of high dark count rate (DCR) and crosstalk of SiPM that seriously hinder its application in low light level detection. The operating principle and performance of the 400 ps time resolution system are discussed along with the improvement of the peak-to-background ratio (PBR) for bulk trinitrotoluene (TNT) Raman spectrum relative to a commercial Raman spectrometer with charge coupled device (CCD). The fluorescence lifetime for solid TNT and Surface Enhanced Raman Scattering (SERS) spectrum for 10−6 mol/L trace TNT have also been obtained by this system, showing excellent versatility and convenience in spectroscopy measurement.



Index Headings: Detectors; Avalanche photodiodes; Photodetectors; Detection; Spectroscopy; Raman; Fluorescence.