holder

The following is an abstract for the selected article. A PDF download of the full text of this article is available here. Members may download full texts at no charge. Non-members may be charged a small fee for certain articles.


Line Selection and Parameter Optimization for Trace Analysis of Uranium in Glass Matrices by Laser-Induced Breakdown Spectroscopy (LIBS)

Volume 67, Number 11 (Nov. 2013) Page 1275-1284

INHEE CHOI, GEORGE C.-Y. CHAN, XIANGLEI MAO, DALE L. PERRY, and RICHARD E. RUSSO*


Laser-induced breakdown spectroscopy (LIBS) has been evaluated for the determination of uranium in real-world samples such as uraninite. NIST Standard Reference Materials were used to evaluate the spectral interferences on detection of uranium. The study addresses the detection limit of LIBS for several uranium lines and their relationship to non-uranium lines, with emphasis on spectral interferences. The data are discussed in the context of optimizing the choice of emission lines for both qualitative and quantitative analyses from a complex spectrum of uranium in the presence of other elements. Temporally resolved spectral emission intensities, line width, and line shifts were characterized to demonstrate the parameter influence on these measurements. The measured uranium line width demonstrates that LIBS acquired with moderately high spectral resolution (e.g., by a 1.25 m spectrometer with a 2400 grooves/mm grating) can be utilized for isotope shift measurements in air at atmospheric pressure with single to tens of parts per million (ppm) level detection limits, as long as an appropriate transition is chosen for analysis.



Index Headings: Laser-induced breakdown spectroscopy; LIBS; Laser ablation; Uranium; Spectral interference.