Intracavity Atomic Absorption with an Inductively Coupled Plasma Atomizer

Volume 38, Number 6 (Dec. 1984) Page 876-879

Downey, S.W.; Nogar, N.S.


An inductively coupled plasma is placed inside the cavity of a flashlamp-pumped dye laser and atomic absorption is measured with the use of an optogalvanic detector. An enhancement factor of 170 is shown for intracavity absorption relative to single-pass, extracavity absorption. Both atom and ion lines can be monitored by this technique. Detection limits of 1 μg/mL for sodium and 0.1 μg/mL for barium were observed. The major noise source is pulse-to-pulse fluctuation in the laser output.