Multisample System for Raman Difference Spectroscopy

Volume 40, Number 4 (June 1986) Page 553-555

Savoie, Rodrigue; Alex, Serge; Langlais, Marc


A new type of sample holder suitable for Raman difference spectroscopy is described. It can accommodate up to six different samples (~10 μL) contained in capillary cells such as those commonly used in conventional Raman spectroscopy. This system can be readily incorporated in an instrument equipped with a conventional computerized control system and it provides a differential frequency accuracy (±0.03 cm−1) which is comparable to that of other existing devices.