holder

The following is an abstract for the selected article. A PDF download of the full text of this article is available here. Members may download full texts at no charge. Non-members may be charged a small fee for certain articles.


Effect on Noise of Intensity-Axis Correction of Spectra Recorded with Charge-Coupled Device Detectors

Volume 56, Number 5 (May 2002) Page 564-569

Jones, Jay Pendell; Wambles, Ronald E.; Mann, Charles K.; Vickers, Thomas J.


We have examined the effect of intensity-axis correction on noise in white light spectra recorded with charge-coupled device (CCD) detectors. Measurements were made with five detectors in Raman spectrometers. Detectors were both liquid-nitrogen and thermoelectrically cooled devices and one room temperature device. Both random and pattern noise have been considered. We used cross-correlation of noise sets to provide an indicator of a fixed pattern in the spectra and an assessment of the efficacy of the correction procedure in removing this pattern. For four of the five detectors intensity-axis correction provided a significant improvement in signal-to-noise ratio. Correction was particularly important for measurements made with lower-cost CCD detectors of the kind proposed for process control instruments.