Comparison of 90 and Epi-geometry for Two-Photon Excitation in Capillaries

Volume 54, Number 10 (Oct. 2000) Page 1490-1494

Zugel, S.A.; Lytle, F.E.


Two geometries of excitation/emission optics are compared for two-photon excitation of time-resolved fluorescence in capillaries. The test fluorophore is the enzymatic tag β-naphthylamine (BNA). Two 0.25 NA (numerical aperture) microscope objectives are used in the 90° geometry: This configuration has a sensitivity of 1.9 X 1010 count M-1 min-1 and a detection limit of 2.5 nM [signal-to-noise (S/N) = 3]. The epi-geometry uses single objectives of 0.25, 0.50, or 1.25 NA. This second configuration has a sensitivity of 7.6 X 1010 count M-1 min-1 and a detection limit of 1.7 nM (S/N = 3) for the 1.25 NA objective. The similar performance of the two configurations is in stark contrast to the factor of 46 enhancement predicted by geometric optics for epi-excitation. The discrepancy is attributed to large increases in collected background counts, which make detection of the signal difficult.