holder

The following is an abstract for the selected article. A PDF download of the full text of this article is available here. Members may download full texts at no charge. Non-members may be charged a small fee for certain articles.


Temperature-Insensitive Near-Infrared Spectroscopic Measurement of Glucose in Aqueous Solutions

Volume 48, Number 4 (April 1994) Page 477-483

Hazen, Kevin H.; Arnold, Mark A.; Small, Gary W.


A digital Fourier filter is combined with partial least-squares (PLS) regression to generate a calibration model for glucose that is insensitive to sample temperature. This model is initially created by using spectra collected over the 5000 to 4000 cm-1 spectral range with samples maintained at 37°C. The analytical utility of the model is evaluated by judging the ability to determine glucose concentrations from a set of prediction spectra. Absorption spectra in this prediction set are obtained by ratioing single-beam spectra collected from solutions at temperatures ranging from 32 to 41°C to reference spectra collected at 37°C. The temperature sensitivity of the underlying water absorption bands creates large baseline variations in prediction spectra that are effectively eliminated by the Fourier filtering step. The best model provides a mean standard error of prediction across temperatures of 0.14 mM (2.52 mg/dL). The benefits of the Fourier filtering step are established, and critical experimental parameters, such as number of PLS factors, mean and standard deviation for the Gaussian shaped Fourier filter, and spectral range, are considered.